User manual

142
Vibration Meter
1
Vibration Level Result 1
All vibration levels are measured and logged simultaneous-
ly. You select which vibration levels should be displayed.
Change Parameter
Turn the rotary wheel •
to select the parameter Lxx.
Press enter •
to open the selection menu and choose
the wideband vibration level to be displayed.
Turn the rotary wheel •
to select the required test result
parameters and confirm with enter
.
Change Font Size
Turn the rotary wheel •
to select the actual test result.
Press enter •
1x, 2x or 3x to set the font size to small,
medium or large.
The XL2 shows 2 vibration levels on the display if a large
font is chosen, otherwise it shows 4 vibration levels.
2
Vibration Level Result 2
Follow the setting instructions for vibration level result 1.
3
Filter
Select one of the following frequency filters:
FLAT no filter applied
1-80Hz Filter 1 - 80 Hz with decay rate of 12 dB /
octave in accordance with DIN 45669-1
1-315Hz Filter 1 - 315 Hz with decay rate of 12 dB /
octave in accordance with DIN 45669-1
10-1000Hz Filter 10 - 1000 Hz with decay rate of 18 dB /
octave in accordance with ISO 2954
4
Result Symbols / Capture
This field offers two functions:
Symbols for measurement results •
15
Upper parameter displayed as line.
Lower parameter displayed as bargraph.