User manual
105
1/12 Octave + Tolerance
Stop Measurement and Data Saving
Press stop • .
The XL2 stores the measurement data automatically.
Press enter • to confirm. The measurement data is stored
on the SD Card in ASCII format.
The measurement is completed.
Audio Recording
The XL2 Analyzer may record a linear audio file (48 kHz, 24 Bit)
of the measurement period together with the spectral data. Just
load a txt-file named “12audio.txt” in the root directory of the
XL2. This requires the installed option Extend Acoustic Pack.
Capture References and Create Tolerances
The Spectral Limits Option extends the function range of the
XL2 with trace capturing, relative curve display and compre-
hensive tolerance handling for the FFT Analysis and the high
resolution 1/12 Oct + Tol spectral analysis.
Features:
Capturing of multiple readings into the internal memory •
Comparing measurement results against captures with rela-•
tive or absolute curve display
Comprehensive tolerance handling •
Creating tolerance masks based on captures for passed / •
failed measurements
Export and import of tolerance and capture files•
True peak level in 1/1 and 1/3 octave resolution•
High-resolution Zoom-FFT up to 0.4 Hz steps in the frequen-•
cy range 5 Hz - 20 kHz
Read the detailed description in the separate chapter
Spectral Limits Option (Capture + Tolerances).