Manual
Signal Balance Error
Indication range 0.0 % to 100 %
Deviation from perfect balance in % or *1
Sweep
Frequency Sweep: Level as function of
frequency.
Time Sweep: Measurement of level, THD+N
and frequency as function of time.
1/3
rd
Octave
Spectrum acc. IEC 1260, class II and ANSI
S1.11-1976, class II from 50 Hz to 20 kHz,
Bargraph for Level
RMS
20 Hz to 20 kHz
Scope
Auto triggering, auto ranging, auto scaling
Filters
Linear, A-weighting, C-weigting, C-message,
Highpass 22 Hz / 60 Hz / 400 Hz, X-Curve
-1
,
Voice bandpass
Input Connectors
XLR balanced, RCA unbalanced
Input Impedance
40 kOhm balanced, 20 kOhm unbalanced
Input RMS
1
( upper meas. limit )
balanced+20 dBu (7.75 V
RMS
) unbalanced
+14 dBu (3.8 V
RMS
)
Max. DC Input
± 50 V
DC
Residual Noise
< 12 µV, XLR-input shorted
Microphone Input (for Polarity measurement only)
Omnidirectional
Monitor Output
Jack 3.5 mm (1/8“), suitable for all common
headsets
Display
Graphic LCD 64 x 100 pixel, with backlight
Batteries
3x AA package dry batteries (alkaline) Typical
battery lifetime > 16 hrs
Dimensions (L x W x H)
163 x 86 x 42 mm (6.4“ x 3.38“ x 1.63“)
Weight
300 g (10.5 oz) incl. batteries
Temperature
0° to +45° C (32° to 113° F)
Humidity
< 90 % R.H., non condensing
Technical Specification
6-02
1
for input levels > 20 dBu (balanced) the ML1 Adapter -20 dB is
available
6.6.
6.6.
6.
TECHNICAL SPECIFICATIONTECHNICAL SPECIFICATION
TECHNICAL SPECIFICATIONTECHNICAL SPECIFICATION
TECHNICAL SPECIFICATION
6.16.1
6.16.1
6.1
Technical Data General FunctionsTechnical Data General Functions
Technical Data General FunctionsTechnical Data General Functions
Technical Data General Functions
Measurements
- Frequency
- Level-RMS, Level-Relative
- THD+N
- vu+PPM
- Polarity Test
- Signal Balance Error
- Sweep, Frequency Sweep, Time Sweep
- 1/3
rd
Octave Spectrum
- Scope
Frequency
Range 10 Hz to 20 kHz
Resolution 4 digits
Accuracy < ± 0.1 %
Level
Units dBu, dBV, V
RMS
Resolution 3 digits (dB-scale) or 4 digits (V-scale)
Accuracy ± 0.5 % @ 1 kHz
Bandwidth 20 Hz to 20 kHz
Flatness ± 0.1 dB
THD+N (Total Harmonic Distortion + Noise)
Meas. Bandwidth 10 Hz to 20 kHz
Resolution 3 digits (dB-scale) or 4 digits (%-scale)
Residual THD+N balanced < -85 dB @ -10 dBu to +20 dBu
unbalanced < -74 dB @ 0 dBu to +14 dBu
vu+PPM (vu-Indicator and Peak Program Meter)
according to IEC 60268 and DIN 45406.
PPM Type I, IIa and Nordic.
Both meters with adjustable reference and
with analog & numerical peak-hold readout.
Polarity Test (with MR1 test signal)
Positive / Negative detection through internal
microphone or XLR/RCA connector. Checks
polarity of tweeters, midrange-speakers,
woofers, sub-woofers and cables down to 10
dB S/N input signal.
Technical Specification
6-01