Manual

21
Start of measurement: The sound pressure level measurement is
re-started every time
The LEVEL SPL screen is entered
The lter selection is changed.
The clock is reset.
The time weighting selection is changed.
The bargraph range selection (RNGE) is changed.
SoundPressureLevel: The equivalent continuous sound pressure
level (also time-averaged sound level) is indicated in dBL
eq
, dBL
Aeq
,
dBL
Ceq
, depending on the selected lter in the lter menu.
Example:
select in the main menu the lter A-WTD
the measurement unit is changing to dBL
Aeq
and the minimum,
maximum and actual sound pressure level readings below are
A-weighted.
Below the time-averaged SPL reading one of the following values
may be displayed (select with the cursor keys):
ACT: actual sound pressure level (SPL)
MIN: minimum sound pressure level acquired during the
measuring time indicated by the clock. The minimum value is
stored until a new lower reading has been taken.
MAX: maximum sound pressure level acquired during the
measuring time indicated by the clock. The maximum value is
stored until a new higher reading has been taken.
Pause: The L
eq
and min./max. measurement may be freezed for a
user dened time.
Select the pause symbol and press the enter key.
The pause symbol is ashing.
The actual SPL reading is continuously displayed and not
freezed.
Select the pause symbol and press the enter key to continue the
measurement, visible by the continuing counting clock.
Clock:Duration in hours:minutes:seconds of the actual measurement;
select the clock and press the enter key to reset the time and restart
the measurement.
Measurement Functions