20
SoundPressureLevel
In the LEVEL SPL mode the Minilyzer measures the integrated-
averaging level over time, the actual, minimum and maximum sound
pressure level.
NOTES • For this function a self powered microphone is
required. NTiAudio does recommend to use the
accessoryMiniSPL(fordetailsseeAccessories).
• Prior the rst measurement the Minilyzer has to
be calibrated to the microphone specications.
As default the ML1 is calibrated according the
MiniSPL.
• NophantompowerisavailableontheXLR-input.
a.SoundPressureLevelPanel
Fig 3.3
Sound Pressure Level Panel
Calibration
Minimum,
Maximum or
Actual SPL
Pause
Clock
Equivalent
Continious
Sound
Pressure
Level
Time
Weighting
SPL Bargraph with manual ranging
To enter the LEVEL SPL mode (Fig 3.3), select LEVEL -> SPL in the
measurement functions submenu.
For most of the ML1 applications, the sound pressure level
measurements require the use of the A-weighting lter. To prevent
wrong lter settings, causing wrong non-usable test results, the default
lter setting of the “LEVEL->SPL” mode is A-weighting (A-WTD).
Note: Each time the LEVEL SPL mode is selected the
A-weightinglterispre-setbydefault.
Measurement Functions