Method 2
Alternatively, you may calibrate entering the group 2 Parameters (see
Section 13.2).
Press [Tare] when “S6 CAL” is displayed.
Display will show “UnLoAd”.
Remove all weights and press [Tare] when the scale is stable.
Display will show “Ld xx”. Place the suggested calibration mass on the
pan.
Press the [Tare] key when the stable indicator is on.
The scale will calibrate to the mass and then return to the parameter
S6 CAL
Press the [Zero] key to return to weighing.
Remove the calibration weight as soon as calibration is complete.
NOTE: If an error message FAIL Hor FAIL Lis shown during calibration, re-
check the calibration and repeat, if necessary. If the error cannot be
corrected contact your dealer or Adam Equipment for advice.