press [Enter] to start the liquid weighing. The weight will be shown in
the last selected unit. After completion of the liquid weighing, the
balance will compute the density of the sample and display it as
“DENSITY XXXX g/cc”
Remove the item from the pan
Press [Mode] to continue with a new sample or press [Esc] to return
to normal weighing
Pressing [Print] will print the density value in g/cc.
10.0 CALIBRATION
The PW series can only be calibrated using either an internal mass (default
method) or an external mass (if setup by the user). See the User Parameters
section.
10.1 MANUAL CALIBRATION
Pressing the [Cal] key will start calibration. Calibration can also be initiated by
a change in internal temperature or a set time period as determined by the
user (see section 13.5).
10.1.1 Calibration using Internal Calibration mass
Check the display is at zero. Tare if necessary
Calibration will begin automatically. When calibration is complete the
balance will return to weighing.
10.1.2 Calibration using External Calibration mass
Pressing [Esc] will abort the calibration at any time
Press the [Cal] key
The display will show the balance setting a new Zero condition by
showing “LOAD 0”. Make sure the pan is empty then press the [Enter]
key to continue
The display will show the busy symbol and then will ask for the value of
the calibration mass to be entered sounding a beep and displaying the
default calibration mass. For example, “CAL MASS 100,000 g” where