9.4.2 Percent Weighing
Percent weighing will be done by defining a certain weight to be 100%. The
weight to be used can either be entered by the user or taken from a sample
Press [Mode] and then the [Up] or [Down] key to select Percent
weighing, “PERCENT” will be displayed
Press [Enter] to enter the function
Display will show, “PERCENT SAMPLE”
Press [Enter] to select the sample method or skip to manually enter the
sample weight as given below
When “LOAD 100 %“ is shown, add the sample
Press [Enter] to set this weight to be 100%. The display will show “REF
WT” and the sample weight in the last selected unit. After a while
“100%” will be displayed. A message “PERCENT” will be prompted on
the top of the display to indicate that the balance is in the percent
weighing mode
Remove the sample and “PERCENT 0.00%” will be displayed
Place an unknown sample to display the percentage weight
To set another weight as 100%, press [Mode] and continue as before
To manually enter a value to be set as 100%, press [Up] or [Down] key
when “PERCENT SAMPLE” is shown to select “PERCENT Ent Wt”
Press [Enter]. Display will show the value in the unit, last used in the
weighing mode
Enter the weight using the numeric entry method (see section 7.1).
Place unknown sample to display the percentage weight
To perform percent weighing with another sample press [Mode] and
continue as before
To return to normal weighing, press [Esc]