From measurement mode:
1. Press the MODE key to select pH
measurement mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P3.0.
See figure
4. Press the ENTER key twice to select
parameter P3.2.
See figure
5. Press the and keys to select 2, 3, 4,
or 5 point pH calibration.
6. Press the ENTER key to confirm selection
and to return to the subgroup menu.
Press the CAL/MEAS key to return to
measurement mode.
pH
SETUP
SETUP
C
C
D
D
P3.2: Selecting number of pH calibration points
Program P3.2 lets you select the number of pH calibration points the meter will use
in calibration mode: 2, 3, 4, or 5. The meter will automatically exit calibration mode
after you have calibrated to your selected number of points.
P3.3 Selecting °C or °F
This meter lets you select between °C and °F units for temperature readings.
From measurement mode:
1. Press the MODE key to select pH
measurement mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P3.0.
See figure
4. Press the ENTER key three times to
select parameter P3.3.
See figure
5. Press the and keys to toggle
between °C and °F.
6. Press the ENTER key to confirm selec-
tion and to return to the subgroup menu.
Press the CAL/MEAS key to return to
measurement mode.
pH
SETUP
E
E
F
NOTES:
You can also switch between °C and °F in SETUP Program P5.2 on page 44 in
% saturation mode or Program P4.3 on page 53 in mg/l (ppm) mode. If you switch
between °C and °F in pH mode, the meter will also switch in % saturation and mg/l
(ppm) modes.
C
SETUP
F
SETUP
F