Manual
4544
13. Accessories
Replacement meters
35618-02 Replacement waterproof pH 300 meter
35618-70 Waterproof pH 300 meter kit. Includes “All-in-One” probe, five each of pH
“Singles” buffer pouches pH 4.01; 7.00; 10.01; and rinse water, one squeeze bottle for
deionized rinse water, batteries, instructions and two electrode brackets,
all in a hard shell carrying case
35618-12 Replacement pH 310 meter
35618-72 Waterproof pH 310 meter kit. Includes “All-in-One” probe, five each of pH
“Singles” buffer pouches pH 4.01; 7.00; 10.01; and rinse water, one squeeze bottle for
deionized rinse water, batteries, instructions and two electrode brackets, all in a hard
shell carrying case
Electrodes and probes
35618-05 Temperature probe. For pH 300 and pH 310 meters
35808-71 Replacement “All-in-One” combination pH/temperature probe,
single junction, sealed, 145 mm L x 12 mm diameter (0.5 lb)
35801-00 pH electrode, epoxy body, single-junction,
145 mm L x 12 mm diameter (0.5 lb)
35805-13 ORP electrode, epoxy body, single-junction,
145 mm L x 12 mm diameter (0.5 lb)
00653-04 Electrode storage solution, 1 pint bottle. Keeps electrode bulb moist for
faster, more accurate readings. (1.1 lbs)
00653-06 Electrode cleaning solution, 1 pint bottle. Removes buildup from
electrodes and maintains electrode sensitivity. (1.1 lbs)
NOTE: Remember to check the temperature calibration when replacing the ATC
probe or when using with an OAKTON
®
"All-in-One" combined pH and
temperature probe. See section 4.5, "Temperature calibration" (page 17).
Continued on next page
12. Specifications
Mode pH Temperature mV
–10.0 to 110.0˚C
-2000 to 2000 mV
Range -2.00 to 16.00 pH (also 14.0 to 230.0°F
(same for Rel mV)
for pH 310 meter)
Resolution 0.01 pH
0.1˚C (also 0.1°F 0.1 mV between ±199.9 mV
for pH 310 meter) 1 mV beyond ±199.9 mV
Accuracy ±0.01 pH
±0.5˚C (also ±0.5˚F ±0.2 mV between ±199.9 mV
for pH 310 meter) ±2 mV beyond ±199.9 mV
Input BNC connector 6-pin plug BNC connector
Calibration
pH 300 meter: Up to five points (pH1.68, 4.01, 7.00, 10.01, 12.45)
pH 310 meter: Up to six points—select from three different buffer sets:
USAbuffers (pH 1.68, 4.01, 7.00, 10.01, 12.45)
NIST buffers (pH 1.68, 4.01, 6.86, 9.18, 12.45)
DIN buffers (pH 1.09, 3.06, 4.65, 6.79, 9.23, 12.75)
Temperature: Offset in 0.1°C increments
mV: Offset up to ±150 mV
Memory:
pH 300 meter: up to 16 data sets (pH or mV plus temperature)
pH 310 meter: up to 50 data sets (pH or mV plus temperature, date, and time)
pH slope range: 80% to 110%
Temperature compensation: automatic (ATC) or manual from 0 to 100°C
Operating temperature: 0 to 50°C
Power: four 1.5 V AAAbatteries (included)
Battery life: > 100 hours
Housing: meets IP67 standards for waterproof and dustproof housing
Dimensions:
Meter: 7.5"L x 3.5"W x 1.75"H
(19.1 cm x 8.9 cm x 4.5 cm)
Boxed: 9.2"L x 8.5"W x 2.75"H
(23.3 cm x 21.6 cm x 7 cm)
Shipping weight: 1.3 lbs (0.6 kg)
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682