Manual

P3.1: Electrode offset
From measurement mode
1. Press the Setup key to enter Set Up
mode.
2. Press the and keys to scroll
through subgroups until you view
parameter P3.0.
See figure
3. Press the ENTER key to select
parameter 3.1.
4. The display shows the electrode offset
value.
If you are in pH measurement mode,
the display shows the mV offset at
pH 7.00. If you have not calibrated at
any buffer, the primary display shows
0.0 mV.
See figure
If you are in mV measurement mode,
the display shows the relative mV
offset. You can adjust the relative mV
offset in relative mV calibration
mode—see page 16 for instructions.
See figure
5. Press the ENTER key to to proceed to
step 4 of P3.2. Or, press the CAL/MEAS
key to return to measurement mode.
3130
P3.0: Viewing electrode data
Program 3 has two “view only”options that lets you check the electrode parameters
for diagnostic purposes.
pH
SETUP
A
A
pH
mV
SETUP
B
B
pH
SETUP
D
D
pH
%
SETUP
E
E
8.4
P3.2: Electrode slope
NOTE: this parameter is available in
pH measurement mode only.
From measurement mode
1. If necessary, press the MODE key to
select pH measurement mode. Press
the Setup key to enter Set Up mode.
2. Press the and keys to scroll
through subgroups until you view
parameter P3.0.
See figure
3. Press the ENTER key twice to select
parameter 3.2.
4. The display shows electrode slope
in percentage. Slope displayed is
the average slope based on the pH
calibrations. Default setting is 100.0.
See figure
5. Press the ENTER key to return to the
subgroup menu. Press the CAL/MEAS
key to return to measurement mode.
R.mV
SETUP
C
C
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682