Manual
1918
IMPORTANT: For manual compensation,
you must disconnect the temperature probe
(see page 9).
1. Switch the meter on. Press the MODE
key to select pH mode.
2. Press the CAL/MEAS key to enter pH
calibration mode. The CAL indicator will
appear above the primary display.
3. While in pH calibration mode, press
the MODE key to enter temperature
calibration mode. The primary display
and secondary display show the last set
value.
See figure
4. Check the temperature of your sample
using an accurate thermometer.
5. Press the ▲ or ▼ keys to offset the
temperature to the measured value
from step 4.
See figure
6. Press ENTER to confirm the selected
temperature and to return to the pH
measurement mode.
The meter will now compensate pH
readings for the manually set temperature.
Notes
To exit this program without confirming the
manual temperature compensation value,
DO NOT press ENTER in step 6. Press
CAL/MEAS instead.
Manual Temperature Compensation
CAL
C
A
A
5.2
5. Measurement
This meter is capable of taking measurements with automatic or manual temperature
compensation. Automatic temperature compensation only occurs when a temperature sensor
is plugged into the meter. If there is no temperature sensor plugged into the meter, the default
manual temperature setting is automatically 25°C. You can manually set the temperature to
match your working conditions using a separate thermometer.
Automatic Temperature Compensation
For automatic temperature compensation
(ATC) simply plug the temperature probe
into the meter (see page 9 for directions).
The ATC indicator will light on the LCD.
See figure
NOTE: If you are using a temperature
probe, the probe must be submersed in the
liquid you are measuring.
pH
ATC
MEAS