User Manual

Page 21 of 44
9.4.4.1 Animal Weighing Setup Parameters
Press [Mode] and then [Up] or [Down] to select Animal Weighing.
When “ANIMAL” is displayed press [Enter] to enter the function
Press the [Up] or [Down] key to select “rUN” or “SEtUP”
Press [Enter] to select “SEtUP” to set up the balance for animal
weighing
Use the [Up] or [Down] key to scroll through the options for setting up
the mode. The display will show “MODE AUTO” or “MODE MANUAL”.
Select “MANUAL” or “AUTO”
Test time
After [Enter] is pressed to select “MODE MANUAL” or “MODE AUTO”,
it will display “TEST TIME XX”, where XX is the number of seconds
over which the balance will average to compute the final weight as set
during last operation
The XX can be changed from 1-200 using the numeric entry method
(see section 7.1)
To confirm the desired value, press [Enter]
Delay
Next, the display will show “DELAY XX” where XX is the time in
seconds taken by the balance before the sampling starts
The XX can be changed from 0-200 using the numeric entry method
(see section 7.1)
To confirm the desired value, press [Enter]
Threshold (During AUTO mode only)
While in the AUTO mode, the display will next show “THRESHOLD XX
g” where XX is the minimum weight in gram of the item required by the
balance to start the process for animal weighing.
The XX can be changed from 1 to 99 using the numeric entry method
(see section 7.1)
To confirm the desired value, press [Enter]
© Adam Equipment Company 2007