User Manual

Page 17 of 44
9.4.2 Percent Weighing
Percent weighing will be done by defining a certain weight to be 100%. The
weight to be used can either be entered by the user or taken from a sample
Press [Mode] and then the [Up] or [Down] key to select Percent
weighing, “PERCENT” will be displayed
Press [Enter] to enter the function
Display will show, “PERCENT SAMPLE”
Press [Enter] to select the sample method
When “LOAD 100 %“ is shown, add the sample
Press [Enter] to set this weight to be 100%, the busy symbol will turn
on. When ready the display will show “100%”
Remove the sample and place an unknown sample to display the
percentage weight
To set another weight as 100%, press [Mode] and continue as before
To manually enter a value to be set as 100%, press [Up] or [Down] key
when “PERCENT SAMPLE” is shown to select “PERCENT Ent Wt”
Press [Enter] to select the manually entered weight method
Enter the weight using the numeric entry method (see section 7.1)
Place unknown sample to display the percentage weight
To perform percent weighing with another sample press [Mode] and
continue as before
To return to normal weighing, press [Esc]
Note: Percentage will be displayed to the maximum number of decimal
places based on the resolution of the balance. To increase or decrease by
one decimal place, press the [Up] or [Down] key respectively.
© Adam Equipment Company 2007