Manual
3.3 LEVELLINGTHE BALANCE
After placing the balance in a suitable place, level it
byusingthespiritlevelontherearofthebalance.To
levelthebalanceturnthetwoadjustablefeetatthe
rearofthebalanceuntilthebubbleinthespiritlevel
iscentered.
3.4 WARM‐UPTIME
Attach the power supply cable to the connector on
the rear of the balance. Plug the power supply
moduleintothemains.Thedisplaywillindicatethe
balance serial number (if set) and the software
revision number followed by the capacity of the
balance. Next the balance will run a self‐test by
displaying all segments followed by a symbol
indicatingthebalanceisinbusymode.Ifthebalance
serialnumberisnotsetthedisplaywillshowdashes.
The display will show zeroes accompanied by the
Î0Ísymbol.
Before you start weighing, you have to wait for the balance to achieve a stable
internaltemperature.
A stable sign ~ is shown when the balance is in
stablecondition.Itwillturnoffifthebalanceisnot
stable.
Exactzeroisshownwhenthe“Î0Í “symbolison
totheleftofthedisplayarea.
7 | Page © Adam Equipment Company 2008