10.0 BATTERYOPERATION
Thescalemayoperatefromabuiltinleadacidbattery.Duringoperationwiththe
batterythebacklightshould bedisabledforbestbatterylife.Typicalbatterylife20
hours.
If possible disable the backlight automatically if the external power supply is not
turnedon.
11.0 ERRORCHECKING
Duringweighingthebalanceisconstantlycheckingtoseeifthebalanceisoperating
withinthelimitedparameters.Theerrorslikelytooccurare:
A/Dcountsbelowlowestallowedvalue
A/Dcountsabovehighestallowedvalue
A/Dnotoperating
Maximumcapacityexceeded
Othererrorsmaybedetectedduringspecialfunctionsoroperations.Thesewillbe
describedinthesectionthatapplies.
Errormessagesandthereasonsare:
Concerningcalibration
ERRORStAb Calibrationcouldnotbecompleted
becausetheresultswerenotstable
ERRORLOWorERRORHIGH Calibrationconstantnotwithin20%ofoldcalibration
constant
Concerningweighing
ERRORLOW Weightdisplayisbelowzeroby>4%max
ERRORHIGH Weightisabovemaximumplus90d
ERRORZERO Weight on pan exceeds zero limit when power is
applied.
29 | Page © Adam Equipment Company 2008