7) Placethemassonthebalance.Press[Enter]to continue
8) On the PGL 203 and PGL 303 models the display will show the busy symbol
and after calibration is complete it will showLOAd 0 on the display.
Removetheweight.Thescalewillreturntoweighing.
9) Allothermodelswillreturntoweighing assoonascalibrationiscomplete.
8.2 AUTOMATICCALIBRATION
The balance will ask for calibration when the balance has automatic calibration
enabledandtheconditionsoftheautomatic calibrationhavebeenmet.
Thecalibrationwillbecalledforifthetimesincelastcalibrationexceedsapreset
time,typically4hours.
Thebalancewillcallforcalibrationtobecarriedoutbyflashingthe“CAL”symbol
onthedisplay.Assoonasthebalanceiscalibratedthesymbolwillbeturnedoff.
The Auto calibration feature can be enabled, disabled or changed by the user
optionstomeettherequirementsoftheusers.Seesection12.5.
8.3 CALIBRATIONERRORS
Occasionally during calibration an error will be detected. These errors can be
causedby:
z Unstablereadings
z Impropercalibrationweightsbeingused
z Largeshiftsofzerofromthefactorysettings
When an error is found a displayed message will be shown and the calibration
must be done again.If the balance has error messages more than once it is
possiblethemechanicshavebeendamaged.
23 | Page © Adam Equipment Company 2008