8.0 CALIBRATION
ThePGLseriescanonlybecalibratedwithanexternalmass.
8.1 MANUALCALIBRATION
Pressingthe[Cal] keywillstartcalibration.Calibrationcanalsobecalledforaftera
settimeperiodasdeterminedbytheuser(seesection12.5).
CalibrationusingExternalCalibrationmass:
1) Pressing[Esc]willabortthecalibrationatanytime
2) Checkthedisplayisatzero.Tareifnecessary
3) Pressthe[Cal]key
4) The display will show the balance setting a new Zero condition by showing
“LOAD 0”.Makesurethepanisemptythenpressthe[Enter]keytocontinue
5) The balance will then show the value of the calibration mass required
soundingabeep,forexample “LOAD 1000 g”
6) If a different mass is desired press the up or down arrow key to show the
alternatemassespossibleforthescale.
PGL
203
PGL
303
PGL
2002
PGL
3002
PGL
4001
PGL
6001
PGL
8001
PGL
10001
PGL
20001
Mass1 100g 100g 1000g 1000g 2000g 3000g 4000g 5kg 10kg
Mass2 200g 200g 2000g 2000g 3000g 4000g 5000g 10kg 20kg
Mass3 300g3000g 4000g 5000g 6000g
Mass46000g 7000g
Mass58000g