Collect Biolm Sample
Identify the water system to be sampled and an appropriate location from which to obtain a biolm sample.
Large systems may need to be sampled and tested at multiple locations. The recommended minimum area to
swabis10cm
2
. If the surface to be sampled is dry, then pre-moisten the swab in recovery buffer, using the re-
suspension tube. Wipe the swab across the area to be tested.
Avoid generating aerosols when collecting or handling samples!
Transfer the swab containing sample to the re-suspension tube and snap off the handle.
Resuspend the Bacteria
Add2mlofRecoveryBuffertothere-suspensiontubecontainingtheswabsampleandscrewonthelid.Shake
thetubeforatleast20secondsoruntiltheswabhasreleasedthebiolmsampleintotherecoverybuffer.
Add Sample to the Test Strip
Removethetestfromitsfoilwrapping,andplacethetestonaatsurface.
The foil wrapping should not be opened until immediately prior to running the test. If the foil is opened and
thenthetestisnotperformedwithin60minutes,discardthetest.
Beforeuse,thetestshouldhave2palebluelinesacrosstheresultwindow.Ifthesearenotpresent,notifyyour
supplier to replace the test kit. Take the pipette from the foil wrapping.
Placeopenendofthepipetteintothesolutioninthere-suspensiontube,thensqueezeandreleasethetop
bulb. This should draw the sample all the way up the long tube and may place a small amount of sample in the
bottom of the bulb. Avoid getting air bubbles in the tube -- the pipette lling process can be repeated if neces-
sary to remove air bubbles.
Placethepipetteoverthesmallsamplewindowattheendoftheteststrip,andsqueezethepipettebulbto
dispense the correct amount of water onto the test strip.
Record the time.
Test Procedure - Swab Test for Biolms
For how to read and interpret the test results, see page 6.