7.0 CALIBRATION
Calibration is recommended for best accuracy. It is suggested that the
scale be calibrated when first installed and any time the scale is moved
to a different environment.
To ensure best accuracy calibrate before use.
Ensure weights are ready before calibrating, the weight should be of a
known accuracy, for example OIML class M1 or Class 4 of ASTM E617.
Do not calibrate the scale if no calibration weight is available.
Calibration can only be accessed if the Calibration function has been
enabled as described in section 6.0.
7.1 AUTOMATIC
EXTERNAL
CALIBRATION
In the weighing mode, press [Func] key to access calibration mode. The
display will show “CAL”.
Press [Zero] key, an access code (Pin code) is required to continue, The
display will show “Pin” .
To enter the access code press the [Func] key, [Unit] key and[Tare] key
in sequence, then press the [Zero] key to confirm the pin code.
The display will show the weighing unit last used to calibrate the scale,
“S-Unit kg”(or ”S-Unit Lb”) ,press [Unit] key to change to lb’s or kg’s for
the calibration weighing unit value.
Press [Zero] key to start the zero calibration. The display will show “CAL
0” and when the stable symbol is displayed press [Zero] key to confirm.
The display will now show the calibration mass expected. ”CAL XX Kg “
or ”CAL XX Lb “ .Press [Tare] key to change the calibration values if
necessary. Each model has 3 different kg and lb calibration selection
values.