7.0 CALIBRATION
TheHighlandseriesofbalance scomesstandardwithHandiCal
TM
internalcalibration
tomakecalibrating thebalancequickandeasy.Howeveryoucanalsocalibratethe
balanceusinganexternalverificationweightifneeded.TheHandiCalmethodisthe
default but if you would like to use external calibration then you must first enable
thisviatheparameter(section6.5SelectingTheInternalOrExternalCalibration).
InternalCalibrationusingHandiCal
1) Pressthe[on/off]keytoturnthepoweron.
2) Press the [Smpl] and [Print] keys at the same time during the self‐checking
test.
3) Thedisplaywillshowunload.Removeanyweightfromthestainlesssteelpan.
4) Whenthestableindicatorshowspressthe[Tare]key.
5) ThedisplayshowsC Int. Pressthe[Tare
]key
6) The display shows LoAd. Lower the internal calibration mass located behind
thepan.Usethehandletohelplowertheweightasfarasitwillgo.Pressthe
[Tare]keyoncethestablesignisshown.
7) The display will show PASS. Rotate the weight back to its off position. Once
removedthebalancewillreturnto zero.
ExternalCalibration
1) Pressthe[on/off]keytoturnthepoweron.
2) Press the [Smpl] and [Print] keys at the same time during the self‐checking
test.
3) Thedisplaywillshowunload.Removeanyweightfromthestainlesssteelpan.
4) Whenthestableindicatorshowspressthe[Tare]key.