Manual
6.6 ADJUSTTHEVALUEOFTHEINTERNALMASS
The internal mass value stored in memory can be adjusted to more closely match
thevalueoftheuser’sexternalmass.
1) To set this parameter pres s the [Mode] key during self‐checking test when
turningonthebalance.
2) Afterafewseconds,thedisplaywillshowthefirstfunctionF1 UNT.
3) Keeppressingthe[Mode]keyuntilF6 CA isdisplayed.
4) Pressthe[Tare]keytoviewthecurrentsettings.
5) Thedisplaywillshowthecurrentvaluewiththefirstdigitflashing.Tochange
the value press [%] to shift the position, press [Mode] to increase the value
andpress[Print]todecreasethevalue.Theinternalcalibrationweightshould
onlychangebetween95.000 and 105.000gramsor495.00and505.00grams
(dependinguponthemodel).
6) Press[Tare]tostoreaparticularsetting.ThedisplaywillreturntoF6 CA
7) When F6 CA isdisplayed press the[Print] key to returnto weighing or press
[Mode]togotothenextfunction.
8) Pressing[Mode]willshowTECH.Thisfunctionisamanufacturingfunctionfor
qualified technicians only and you can press [Mode] again to scroll through
parameteroptions,orpress[Print]toreturntonormalweighing.
Changingthevalueoftheinternalweightwillaffectthecalibrationwhenusingthe
internal calibration. To check if your internal weight is correct you should only use
highqualityweightsthatareaccuratetothereadabilityofthebalance.
19|Page©AdamEquipmentCompany2008