6.4 AUTOPOWEROFF
The auto power off function helps conserve power, when using the internal
rechargeablebatteryorACadapter.TheAutoswitchofftimemaybesetupbythe
user.
1) To set this parameter pres s the [Mode] key during selfchecking test when
turningonthebalance.
2) Afterafewseconds,thedisplaywillshowthefirstfunctionF1 UNT. 
3) Keeppressingthe[Mode]keyuntilF4 oFFisdisplayed.
4) Pressthe[Tare]keytoviewthecurrentsettings.
5) Press[Mode]tochangethesettings(0,5,10,20and30minutes).
6) Press[Tare]tostoreaparticularsetting.ThedisplaywillreturntoF4 oFF
7) WhenF4 oFFisdisplayedpressthe [Print]keytoreturntoweighingorpress
[Mode]togotothenextfunction.
6.5 SELECTING
THEINTERNAL OREXTERNALCALIBRATION
Youcanselectiftheinternalmassoranexternalmassistobeusedtocalibratethe
balance.
1) To set this parameter pres s the [Mode] key during selfchecking test when
turningonthebalance.
2) Afterafewseconds,thedisplaywillshowthefirstfunctionF1 UNT.
3) Keeppressingthe[Mode]keyuntilF5 IEC isdisplayed.
4) Pressthe[Tare]keytoviewthecurrentsettings.
5) To change the settings press the [Mode] key to change from INt (internal
calibration)orE(externalcalibration)
6) Press[Tare]tostoreaparticularsetting.ThedisplaywillreturntoF5 IEC
7) WhenF5 IEC is displayedpress the[Print]key toreturntoweighingorpress
[Mode]togotothenextfunction.
18|Page©AdamEquipmentCompany2008