User guide
19
Front panel displays and controls continued
7 AVG button
8 HOLD button
+-
ALARM
EXIT
MENU
1
STORE
2
LOG
3
RECALL
4
PRINT
5
MAX
6
MIN
7
AVG
8
HOLD
9
SCAN
°F
0
CLEAR
.
7
8
7
AVG
7 AVG button
Push to display the average temperature for each channel and the
number of readings that are being averaged. The display continues to
scan channels.
When a numerical entry is required and the NUM icon is on, push to
enter a ‘7’.
•“AVG” appears and remains in the display.
• Readings from when the instrument was last powered up or
cleared are averaged.
• Opened channel readings are not averaged.
• Use Left-Right arrow keys to stop scanning and to choose the
channel whose average is displayed. Push SCAN to return to
scanning channels while remaining in AVG.
• Push AVG again to exit.
8 HOLD button
Push to hold the current record in display.
When a numerical entry is required, and the NUM icon is on, push to
enter an ‘8’.
•“HOLD” appears and remains in the display.
• Channel scanning, logging, printing continue normally but the
display remains.
• Push HOLD or EXIT to exit.
5
°F
AVG
8
HOLD
BA 13-25 10/12/01, 11:03 AM19