User guide

5.0 MEASUREMENT OF SOLID DENSITY
To perform the density of solids test, the user must have a method to
immerse the solid item in a chosen liquid. The density of the liquid must
be known or determined from a look-up table.
For water and ethanol the density will be calculated based on the
temperatures entered using the numeric entry method, when asked by
the balance.
Numeric Entry Method-
To set a value when required, use the keys as given below-
- [Up] and [Down] keys to increase or decrease the flashing digit,
- [Advance] and [Back] keys to advance or move back the digit
and
- [Enter] key to accept the value
Steps:
After installing the Density Determination Kit, tare the balance by
pressing the [Î0/TÍ] key
Press [Mode] and then [Up] and [Down] keys to select “Density”
Press [Enter] to enter the Density mode
Press the [Up] and [Down] keys to select the Solid or Liquid
method when “DENSITY SOLId”” or “DENSITY LIqUId” is
displayed
Press [Enter] to select the “DENSITY SOLId” method
Press [Up] or [Down] to select the liquid of choice- Water, Ethanol
or Other
Press [Enter] to select the choice. For water and Ethanol the
temperature will be asked for. Enter the temperature using the
numeric entry method
For the “Other” choice the density will be asked for. Enter the
© Adam Equipment Company 2007
4