Adam Equipment DUNE SERIES (P.N.
Easy Reference: Model name of the scale: Serial number of the unit: Software revision number (Displayed when power is first turned on): Date of Purchase: Name of the supplier and place: 2|P a g e © Adam Equipment Company 2008
1.0 CONTENTS 1.0 CONTENTS..................................................................................... 3 2.0 INTRODUCTION ............................................................................. 3 3.0 SET UP ........................................................................................... 4 3.1 UNPACKING AND SETTING UP YOUR SCALE ............................... 4 3.2 LOCATING AND PROTECTING YOUR SCALE................................. 5 4.0 BASIC OPERATION .............................
3.0 SET UP 3.1 UNPACKING AND SETTING UP YOUR SCALE 1) Remove the top pan and card board packaging between top pan and machine. 2) Connect the power adapter to the socket on the back of the machine. Installing Batteries: Remove the battery compartment cover and insert six batteries (R6P/LR6 /AA size) into the battery compartment taking extreme care that the polarities plus and minus are observed.
3.2 LOCATING AND PROTECTING YOUR SCALE In order to keep you scale functioning at its best we suggest that you do the following: Avoid extremes of temperature. Do not place in direct sunlight or near air conditioning vents. Make sure the scale is located on a strong table and free from vibration. Avoid unstable power sources. Do not use near large users of electricity such as welding equipment or large motors. Do not mix batteries and use only the factory approved power adapter supplied with the machine.
4.0 BASIC OPERATION 4.1 TURING ON THE SCALE Plug in the unit using the AC adapter or use batteries. Do not use battery power and the AC adapter at the same time. 1) To turn on press the [on/off] key once and release. All of the segments on the display will light up, then the display shows zero >0< 2) The scale is ready to be use. 3) To turn the scale off after use press the [on/off] key again.
5.0 FUNCTIONS 5.1 AUTO POWER OFF The auto power off helps conserve power when using on battery or the AC adapter, the machine will automatically turn off after 4 min of inactive use or no change in the displayed weight. 6.0 CALIBRATION Before calibrating the Dune series it is best to turn on the unit for about 1 min to let the components warm up. 1) Press [on/off] key to turn the power on, if it is not already. 2) Press and hold the [Unit] key, until “CAL” is displayed 3) Remove any weight from the pan.
7.0 TROUBLE SHOOTING 7.1 ERROR MESSAGES If an error message is shown, repeat the step that caused the message. If the error message is still shown then contact your dealer for support. ERROR CODE O-Ld DESCRIPTION POSSIBLE CAUSES Overload OUt2 Out of Zero reading UnSt Unstable LO or Low Battery indicator ErrE E‐Prom error Remove weight from the pan. Make sure the machine is reading zero before weighing. Do not over load the pan. If the problem recurs calibrate. Possible damage to the mechanics.
8.0 SPECIFICATIONS Model # Maximum Capacity Readability Repeatability (Std Dev) Linearity ± Units of Measure Stabilisation Time Operating Temperature Power Supply Calibration Calibration Mass Display Balance Housing Pan Size Overall Dimensions (w x d x h) Gross Weight Features 9|P a g e DCT‐201 200g 0.1g 0.1g 0.
9.0 WARRANTY STATEMENT Adam Equipment offers Limited Warranty (Parts and Labour) for the components failed due to defects in materials or workmanship. Warranty starts from the date of delivery. During the warranty period, should any repairs be necessary, the customer must inform the supplier or Adam Equipment. The company or its authorised Technician reserves the right to repair or replace any components at its own discretion.
Manufacturer’s Declaration of Conformity This product has been manufactured in accordance with the harmonised European standards, following the provisions of the below stated directives: Electro Magnetic Compatibility Directive 2004/108/EN Low Voltage Directive 73/23/EEC Adam Equipment Co. Ltd. Bond Avenue, Denbigh East Milton Keynes, MK1 1SW United Kingdom FCC COMPLIANCE This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.
ADAM EQUIPMENT is an ISO 9001:2000 certified global company with more than 35 years experience in the production and sale of electronic weighing equipment. For a complete listing of all Adam scales and balances visit our website at: www.adamequipment.com © Copyright by Adam Equipment Co. Ltd. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior permission of Adam Equipment.