6.0 CALIBRATION
Calibrationmayberequiredwhenthebalanceisinitiallyinstalledormovedtoanew
location.Overtimeanduse,mechanicaldeviationsmayoccursoitisrecommended
thatyoucalibrateyoubalanceregularlyforpeakperformance.
Beforecalibratingyourshouldmakesureyouhavethecorrectweights.Theweights
shouldbeknowntoanaccuracythatisappropriateforthebalancebeingcalibrate d,
for example, OIML Class M1 type or ASTM E617 Class 4. If you do not have the
correctweightsdonotattemptcalibration.
Model CQT202 CQT251 CQT601 CQT1501 CQT2601 CQT2000 CQT5000
Calibrat
ion
Weight
100g
Or
200g
100g
Or
200g
200g
Or
400g
Or
600g
500g
Or
1000g
Or
1500g
1000g
Or
2000g
1000g
Or
2000g
1000g
Or
3000g
or
5000g
1) Toenterintothecalibrationparameterturnthebalance off,press[Tare]and
[Unit] keys at the same time and switch on the balance after 3 seconds
release[tare]and[unit]keys.
2) ThedisplaywillshowUnLOADremoveanyobjectsfromthepan.
3) Pressthe[Tare]keytoenterazerovalue.
4) The value last used for calibration will be displayed (see above table). To
selectadifferentcalibrationweight,pressthe[Unit]key.
5) Once you have selected the calibration weight to be used press the [Tare]
key.
6) LOADwillbedisplayed,placethecalibrationweightonthepan.
7) Oncethestablesignisshownpressthe[Tare]key.
8) If the weight is within 5% of the last calibration value the display will show
PASS and exit the calibration mode. If the calibration is not successful the
displaywillshowFAILandexitthecalibrationmode,ifthecalibrationfailstry
itagain.