Product Manual
34
vertical index error adjustment mode. The instrument
displays vertical angle on the second line and “1” on
the third line.
(3)Rotate the telescope up and down to initialize the
vertical angle, aim the cross-hairs of collimator or the
target which is not apart away from horizontal line ±
10°, then press and holding down key, the
buzzer sound once, and about 2 seconds later, release
the key, the LCD will display “2” on the third line.
(4) Turn the instrument around 180°, and aim the
same target, press and holding down the
key, the buzzer sound once, and about 2 seconds later, release the key, the
adjustment finish and enter into measurement mode automatically.
NOTE:
1. After the adjustment of vertical index I, please check it again.
2. If the horizontal angle in absolute encoder mode, then it needs initialization to
enter measurement mode.
Vertical index I adjustment
Vz: 091°51′37″
H
R
: 2
Turn 180°and aim the same
target. press and hold down
① key, about 2 seconds
laser, release it, the
instrument come into angle
measurement mode