Product Manual

11
4.3 Centering
4.3.1 Centering with the optical plummet
4.3.1.1 Three-jaw tribrach
Rotate the focusing knob of the optical plummet and
adjust the focus to the ground mark point. Then loosen
the center screw of the tripod. Look through the
optical plummet, and move the base plate on the tripod
head until the center mark coincides with the ground
mark point. Tighten the center screw. Ensuring that the
bubble stays positioned in the center when rotating the
Plate level in steps of 90°(100g). If the bubble is not
positioned in the center, adjust the level screws.
Repeat above steps. Ensure that the center of reticle
plate coincides with the mark point when rotating
alidade of instrument.
Fig. 13
Cross mark
Plummet
center