3-12 Operating Characteristics
3.2.1.2.5 On-Line/REMOTE
The remote Host Computer has control of the GES-1993 using S2F41 (Remote
Command). For details see "Remote Commands” section.
The Operator may issue the PAUSE, STOP, and ABORT commands while in
Remote. These commands will generate an OperatorCommand event (CEID 4) as
well as other messages attached to these activities. Refer to appendix B for further
detail.
3.2.2 Power Up
At Power Up, the Equipment Constant INITCONTROLSTATE (VID 8) controls
whether ControlState is initialized to ONLINE or OFFLINE. Within the Off-Line
super-state, the Equipment Constant OFFLINESUBSTATE (VID 49) controls
whether the Control State defaults to:
1. Equipment Off-Line
2. Host Off-Line
3. Attempt On-Line
If INITCONTROLSTATE (VID 8) initializes to On-Line, the variable
ONLINESUBSTATE (VID 51) determines whether the default state is:
1. Local
2. Remote
When the operator changes the default control state through GEM property page
dialog/Setup page, INITCONTROLSTATE gets updated. ONLINE/REMOTE is
recommended for this value.