System Configuration and Setup 4-85
4.4.7.2 Calibrate
Clicking on Calibrate in the Setup Vision Window opens a Teach Window in which you are prompted to
perform a step-by-step procedure for calibrating the number of pixels in the video display to the number
of X and Y motor steps it takes the dispensing head to travel the equivalent distance.
To calibrate the vision system:
1. Select
Configuration > Setup Vision from the Fluidmove Main Menu.
The Setup Vision window opens (Figure 4-71).
2. Click on
Calibrate.
The Calibration Teach window opens (Figure 4-73).
3. Follow the screen prompts and click
Next to move to the next step.
4. When you are finished click on
Done.
You will return to the Setup Vision window.
NOTE If the camera is replaced or adjusted in any way, the Vision System must be recalibrated.
Figure 4-73 Calibration Teach Window