4.4.6.2 Scale Calibration
The scale is designed to self calibrate to the calibration weight value.
To calibrate the scale:
1. Zero the scale. See 4.4.6 Scale Setup.
2. Click on the
Advanced tab.
Figure 4-70 Scale Setup - Advanced Tab
3. Type in the actual weight of the precision weight that will be used for the calibration. We
recommend using 50 grams as a standard on the Nordson ASYMTEK Dispensing Systems.
4. Set the weight on the scale.
5. Click on
Calibrate.