4-60 System Configuration and Setup
47. Click on Run next to “6. Calculate the Camera-to-Height Sensor XY offset.”
!A Teach Window opens.
!If your system does not have a Height Sensor, this step will be grayed-out and inactive.
48. Choose an easily identifiable mark on the sample substrate, such as a dot or a line intersection,
and use the Position Controls to center the Height Sensor probe tip on the selected mark.
49. Click on
Teach.
50. Use the Position Controls to center the Camera crosshairs over the same mark.
51. Click on
Teach.
52. Click on
Done to accept the results.
!The Teach Window closes and you return to the Setup dialog box. Notice that the box
next to number 6 now has a check mark (9) in it to indicate that Step 6 is complete.
53. If your dispensing system is equipped with a Dual-Action Dispensing Head, click on
Valve 2
in the Valve Offsets dialog box and perform the entire procedure again, except “1. Teach
Safe Z.”
54. When routine is complete, click
End in the Valve Offsets dialog box.