Instruction Manual
Setup and Configuration 6-19
Calculate Master Offsets
This screen-prompted routine identifies and records a Safe-Z Height for dispensing head travel, the XYZ
locations of the purge and weigh stations, calibrates the camera-to-needle offset and the probe-to-needle
offset. CMO also calibrates the camera crosshair to a known location.
This routine is performed when the dispensing system is initially installed and configured. Those values
and settings are recorded in FmNT. You can run this routine again at any time to insure dispensing
accuracy or reflect hardware changes. You can choose to use the pre-recorded values during the routine.
Definitely perform a Calculate Master Offsets if any of the following events occur:
• initial system setup (re-installation of dispensing system)
• re-initialization or re-installation of system software
• software update
• valve change (not syringe)
• camera change or location adjustment
• service station (Purge, Weigh Stations) change or location adjustment
• Needle Sensor change or location adjustment
• Height Sensor change or location adjustment
• changing from one of the other offsets routines
To perform a Calculate Master Offsets procedure:
Follow all on-screen prompts.
1. In the Main Window, click on
Configuration.
2. Click on
Machine Offsets on the Configuration menu.
3. Select
Calculate Master Offsets from the drop-down menu.
?NOTE If Calculate Master Offsets is greyed-out, refer to “Changing Offsets Mode”
later in this section of the manual, then return here.
> A Teach Window opens.
4. Follow all on-screen prompts.
> Refer to the Online Help for assistance.