3-4 Operation
Setting the Fiber Optic Fan Sensor Sensitivity
The following steps describe how to calibrate the sensitivity of the Fiber Optic Sensor Amplifier in the
Windows environment.
?NOTE The Fiber Optic Sensor Amplifier settings should be as follows:
Switches: D.ON
Received Light Intensity: No ā€œPā€ after the number on the display.
Item Description
1 Power Selection Bar Graph
2 Timer Selection Bar Graph
3 Up-Down Button
4 Mode Button
5Switch
6 Set Button
7Display
8LED
Figure 3-2 Sensor Amplifier Settings
1. Remove the clear front window from the Fan Width Enclosure by unscrewing the four
screws.
2. Press the
MODE button on the Sensor Amplifier for 3 seconds until one of the selections
under the Power Selection bar graph starts flashing. See Figure 3-2.
3. Press the Up-Down button on the Sensor Amplifier until the
TURBO light is on.
4. Press the
MODE button again and one of the selections under the Timer Selection bar graph
will flash. See Figure 3-2.
5. Press the Up-Down button until the
40ms light is on.
6. Press the
MODE button again to return to the operational display.
7. Move the dispenser so the spray pattern will be in between the sensors.
8. Press the Up-Down button once and the display will flash.
1
2
3
4
56
7
8