Upgrade Kit Owner manual

Upgrading the DP-3000 7
c. Select Setup Purge Station.
d. Make sure that the checkbox labeled Timed Purge is NOT checked.
e. Click on Purge.
11. Press OK.
12. Using the adjustable wrench, turn the piston tool, and remove the piston.
" NOTE The dispensing system interlocks must be disabled or the doors must be closed when
you click Purge.
- See the DP-3000 Owner’s Manual for a description of DP-3000 commands.
- The piston will fully extend from the pump body and stop at the bottom of
its travel.
- A dialog box opens indicating “DP-3000 out of travel limits”.
Alternately, you can use a 5/8-inch socket and ratchet. Place thread sealant or Teflon
tape around the piston threads to protect the lead screw from fluid leakage into the
receiver block.
CAUTION! Failure to properly seal the piston threads could cause irreparable damage to the
pump if the priming O-ring is not in place, causing fluid to leak into the receiver
block. The fluid could cure around the lead screw, preventing it from movement.
13. Place the piston extraction tool around the new piston, and place both into the pump. Fit the
receiver body around the piston.
14. Using the 2.5-mm hex key, tighten the screws.
15. Install the slide retainer onto the receiver block, and secure it with the receiver pins.
! You will hear an audible click when the pins are correctly installed.
16. Refer to the DP-3000 Owner’s Manual for instructions on reinstalling the pumping chamber.