Instruction Manual

Operation 39
To install the stopcock/needle assembly into the pump:
1. Assemble the stopcock and needle assembly and insert it into the Luer clamp block as shown
in Figure 3-29.
2. Flip the yellow Luer clamp block lever down in the direction shown in Figure 3-30.
! The stopcock/needle assembly should now be held firmly in place.
3. Pull the blue sliding bracket detent lever toward you, withdrawing the rod from the detent.
! This frees the sliding bracket.
4. Move the sliding bracket upward until the stopcock taper meets the pump taper.
5. Rotate the green lock lever down.
! This raises the carriage and locks it in place, forcing the stopcock taper up against the
pump taper to create intimate contact and a good seal.
O
F
F
O
F
F
Figure 3-29 Inserting Stopcock/Needle
Assembly into Pump
Figure 3-30 Securing the Stopcock/Needle Assembly
Support from the
sliding bracket to
the stopcock
Luer Clamp Block
Insert the
stopcock/needle
assembly into
Luer clamp block
(1)
Lock Lever
(green)
Luer clamp block
lever
(y
ellow
)
Sliding bracket
detent lever (blue) in
unlocked position
After pulling the sliding
bracket detent lever out,
push the sliding bracket
upward.
(2)
(3)
(4)
(5)
Detent