Instruction Manual
Operation 35
Needle Sizes and Types
Changing the size of the dispensing needle can change the width of the fluid line dispensed. A larger
needle will create a larger line. However, using a smaller needle does not affect the flow rate. The
DP 3000 Series pump accepts Luer lock dispensing tips. Use only the dispensing tips provided in the
VK 6000 valve kit.
" NOTE You should always use the largest diameter needle possible for your application to
place the least amount of stress on the stopcock. Higher viscosity fluids can destroy the
stopcock if used with a too small needle size.
Preventing Needle Movement
If the stopcock and needle assembly is not assembled correctly, the needle may move during the
dispensing process. This movement does not impact pump placement accuracy, and the movement is
repeatable—the needle always returns to its original position. However, proper assembly of the stopcock
and needle can eliminate needle movement.
There are four contact surfaces on the DP-3000 that provide support for the stopcock and needle
assembly. See Figure 3-24.
1. The side of the sliding bracket supports the back of the stopcock.
2. The bottom surface of the stopcock rests on the upper surface of the Luer clamp block.
3. The Luer clamp ring is clamped into the Luer clamp block.
4. The pump taper on the bottom of the pump chamber is forced into the stopcock taper. This
mating provides a seal as well as a downward force to push the stopcock against the
surfaces mentioned in points 1 and 2 above.
O
F
F
O
F
F
Figure 3-24 Stopcock and Needle Assembly Support
Stopcock and needle assembly
slides down into Luer clamp
block
(3) Stopcock Luer lock
ring inside Luer clamp
block (not visible here)
(2) Support from the
Luer clamp block to
the stopcoc
k
(1) Side of sliding
bracket
(4) Stopcock Taper
(4) Pump Taper