Quick Operations User Manual

36 Operation
System Status Controls
Table 3-4 System Status Controls
Button/Indicator Function
CONVEYOR
STATUS
RESET
V
PAUSE
CONVEYOR
STATUS
RESET
VV
PAUSE
CONVEYOR STATUS has two buttons.
PAUSE halts the Conveyor. The PAUSE LED lights up to indicate that
the Conveyor has been paused. Press PAUSE again to resume
Conveyor movement.
Press RESET to clear the Conveyor Controller after a fault condition
has occurred.
INTERLOCK
INTERLOCK
RECOVERY
The Interlock is a key-activated safety feature that prevents the
Conveyor and the Dispensing Head from moving while the dispensing
system Front Doors are open.
The Interlock should always be ON during normal operation. This helps
prevent personal injury or damage to the dispensing system.
For more information on the Interlock feature, see “
Safety Interlock
System” in the
Safety section.
SYSTEM
STATUS
RESET
SYSTEM
STATUS
RESET
SYSTEM STATUS LEDs are color coded for each System Status
condition. The LED lights correspond with the colored lights on the
Light Beacon. For an explanation of the different status conditions, see
Table 2-2.
If the red ABORT LED has been triggered, press the SYSTEM
STATUS RESET after the condition that triggered it has been
corrected. This will reset the dispensing system electronics.
" NOTE After you press the SYSTEM STATUS RESET
button, you must press the FIND HOME button
to re-establish motor references.
VENT AIR
MAIN AIR
The VENT AIR LED illuminates when vent air is insufficient.
" NOTE The VENT AIR LED flashes while the dispensing
system is powering up.
The MAIN AIR LED illuminates when main (facility) air pressure is
insufficient.
APPLIC
ON/OFF
APPLIC
ON/OFF
Press APPLIC ON/OFF to turn fluid applicator operation on and off.
This feature can be used to test run a program without dispensing fluid
(dry mode).
PURGE
This button is NON-FUNCTIONAL on the C-740/C-741 dispensing
system. The purge function is executed through the software.