User guide

Height Sensor 10-5
Hardware Setup for HS-Series
Once you have installed your HS-Series Height Sensor, you should perform the following hardware setup
procedure to make sure that the Height Sensor is in a neutral position before making any probe
adjustments or performing the configuration procedure. Perform any necessary Probe Adjustments,
following the procedures outlined in “Probe Adjustments for HS-Series” later in this section.
?
NOTE The following procedure may be performed by either an operator or a service technician.
To set up the HS-Series Height Sensor
1. Check the location of the Micrometer Locking Screw (see Figure 10-3).
> The screw should be centered in the slot on the bracket. If the screw is already centered in the
slot, the hardware setup is complete. STOP here. If the screw is not centered in the slot,
continue with the next step.
2. Loosen the Micrometer Locking Screw.
3. Adjust the Micrometer Adjustment until the Micrometer Locking Screw is centered in the slot.
4. After centering the screw, it is recommended that you adjust the Micrometer Adjustment to “0” as
a starting point for any future probe adjustments.
5. Tighten the Micrometer Locking Screw.
6. Perform Probe Adjustments, if necessary.