Troubleshooting 6-5
Testing System Repeatability
To test system repeatability:
1. Use the directional controls on the dispenser Control Panel to move the Dispensing Head to a
particular location within the dispensing area.
2. Press
Set Origin on the dispenser Control Panel.
3. Use the directional controls on the Dispenser Control Panel to move the Dispensing Head to a
different location within the dispensing area.
4. Press
Go To Origin on the dispenser Control Panel.
> The Dispensing Head should move back to the first position within ± 0.001 in (0.0254 mm).
If it does not, call Asymtek Technical Support.