User guide
Special Options 13-13
Operation Controls
Tool Required
• Flathead screwdriver
?
NOTE If you need to change the Main Air pressure, see “Main Air Inlet” in the Operation
section.
To set the low air pressure sensor set point:
1. Set the system pressure with the Valve Pressure or Fluid Pressure regulators to the low kPa (psi)
at which you want to trigger the alarm.
> See “Gauges and Regulators” in the Operation section for instructions on how to set pressure.
2. Slide the collar down on the Low Air Pressure Alarm sensor.
3. Turn Safety Interlock ON (I).
> The Dispenser Status green LED indicator on the Control Panel and on the Light Beacon
will turn ON.
4. Using a flathead screwdriver, turn the brass sensor adjuster counter-clockwise until the sensor is
triggered and the
Dispenser Status yellow LED indicator and light beacon Turn ON.
5. Adjust the Main Air to the desired pressure (see “Main Air Inlet” in the Operation section).
> The Dispenser Status yellow LED indicators and the Light Beacon will turn OFF and the
green LED will turn ON.