Instruction Manual

6-4 Troubleshooting
Table 6-1 System Troubleshooting (continued)
Symptom Possible Cause Recovery
A. System not
connected to
facility air supply.
Check the Main Air Inlet at the rear of the
system.
8. No Air Pressure.
B. Valve air
pneumatic hose
not connected
Check valve air; secure connections if
necessary
9. Air is hissing at the
rear of the system.
Facility air pressure is
too high (>100 psi).
Disconnect Main Air supply and install a
facility regulator.
A. Needle Sensor not
set up correctly.
1. From the FMW Main Window, click on
3 Configuration.
2. Select Calculate Master Offsets.
See Section 5 – Startup.
10. Needle Sensor
cannot find needle.
B. Needle Sensor
beam dirty (light
always on or lags
on)
Clean Needle Sensor with IPA or
compressed dry air.