Installation 3-5
Connecting Power and Air Supply
CAUTION! This procedure should only be performed by a trained service technician.
To connect the facilities power and air supply:
1. Unpack the Main Power cable and remove all packing material including plastic wrappers and
tie wraps.
2. Plug the female end of the power cable into the Main Power Inlet on the back of the
dispensing system. Plug the male end of the power cable into the facility power source.
3. Locate and attach the Quick Disconnect Air Nozzle to the facility air hose.
4. Connect the facility air hose to the Main Air Inlet connector on the back right side of the
dispensing system.
5. Install a damper at the exhaust duct.
6. Connect the Facility Exhaust Ventilation System ductwork to the Exhaust Vent Duct. See
Figure 3-1.
7. Adjust airflow to a minimum of 400 SCFM.
Item Description
1 Main Power Inlet
2 Exhaust Vent Duct
3 Main Air Inlet
Figure 3-1 Connecting the Facilities Power and Air Supply
1
2
3