Owner manual

3-4 Installation
Pneumatic Connections
The C-341 Series Dispensing System requires 85 l/min (3 SCFM) of clean, dry air delivered through a
12.7 mm (0.5 in.) pipe or hose at 90 psi (620 kPa). Connect the air supply to the quick-disconnect fitting
at the main air connection. See
Figure 3-1.
" NOTE The maximum air supply pressure should not exceed 100 psi (689 kPa).
WARNING! All system pneumatic connections should be checked before the main air is
connected.
Electrical Connections
The dispensing system requires 1 kW of single-phase power at 100, 120, 220, or 240 VAC.
" NOTE Robot Controller board wiring connections are installed in the factory before the
product is shipped. Refer to the Specifications section for more information on
electrical requirements. Contact your Asymtek representative for requirements on
your particular system.
Refer to Appendix A for C-341 Series Dispensing System electrical diagram and electronic connection
information.
Exhaust Connection
The dispensing system requires a 15.3 cm (6 in.) diameter exhaust duct capable of drawing a minimum of
0.28 m
3
/s (600 SCFM) air at 25.4 mm (1 in.) water column (w.c.) static pressure.
" NOTE Refer to the Specifications section for more information on exhaust requirements.
Contact your Asymtek representative for requirements on your particular system.