Installation Manual

5-14 Power-up and Testing
Light Beacon
Perform the following procedures to check function of the Light Beacon.
To test the Light Beacon:
1. Close the dispensing system Hatch, make sure both EMOs are deactivated, and press the
green
ON (l) button on the Operator’s Console.
> The green Beacon light should be ON.
2. Open the Hatch.
> The yellow Beacon light should come ON and the green Beacon light should go OFF.
3. Close the dispensing system Hatch.
> The green Beacon light should come ON and the yellow Beacon light should go OFF.
4. In the FmNT Main Window, click on
Tools.
5. In the Tools Window, click on
I/O Test.
6. Click on
Dispenser.
> The Dispenser I/O Test dialog box opens. See Figure 5-8.
7. On the Channels 10-13 tab, locate the bits listed in Table 5-1 and toggle each one ON and
OFF while watching the Light Beacon for the desired result.
Table 5-1 Light Beacon Test Matrix
Bit Name Toggle On Result Toggle Off Result
BEACON_RED Red Light ON Red Light OFF
BEACON_YEL Yellow Light ON Yellow Light OFF
BEACON_BLU Blue Light ON Blue Light OFF
BEACON_REDFLASH Red Light ON Red Light OFF
BEACON_YELFLASH Yellow Light Flashes Yellow Light OFF
BEACON_BLUFLASH Blue Light Flashes Blue Light OFF
ALARM_SOUND Alarm Sounds Alarm OFF
8. Click on OK to return to the Main Window.