Installation Manual
Uncrating and Placing 3-5
Unpacking the Dispensing Chamber
CAUTION! If the dispensing system is being installed in a clean room, remove all packaging
material and follow facility-recommended procedures before moving components
into the clean room.
To unpack the Dispensing Chamber:
1. After all perimeter packing material has been removed from the dispensing system, remove
the keyboard and mouse from their boxes and place them in the keyboard tray.
2. Remove all tie wraps, tape, foam packing material, and warning tags from the following
areas. See Figure 3-3 for packaging of a typical Axiom X-1000 Dispensing System.
• Dispensing Head
• Conveyor
• Dispensing Calibration Module
? NOTE Location of packaging materials to be removed is indicated by the presence of
red warning tags. The amount, type, and arrangement of packaging materials will
depend on your system’s configuration.
3. Loosen the 3-mm setscrews on the X- and Y-axis Dispensing Head Stoppers. See Figure 3-4.
4. Remove the Stoppers and retain them in a safe place.