Installation Manual
8-146 Troubleshooting
10. Scroll down to [SCALE DI104] and confirm the following settings:
• [SCALE DI104] • Name = DI104
• Port Type = Serial • Port Name = Com1
• Baud Rate = 9600 • Parity = None
• Data Bits = 8 • Stop Bits = 1
• Event Char = None • Min Free Space = 0
• RTS-CTS = 0 • DTR = 0
11. Did the [SCALE DI104] file match the settings?
Go to Step 12
Contact Asymtek Technical Support
12. Make corrections to the [SCALE DI104] file, save settings, and test Scale in accordance with Steps 5
through 10 of “Scale Operational Verification” in this subsection.
13. Shut down Windows NT, press the black
OFF (0) button on the Operator’s Console, switch Main
Circuit Breaker to
OFF (0) position, and disconnect Power Cable from the facility outlet.
14. Disconnect the Scale Communications Cable (P/N 06-4582-00) from Com1 and connect it to Com2.
15. Connect the Power Cable to facility outlet, switch the Main Circuit Breaker to
ON (I) position, and
press the green
ON (I) button on the Operator’s Console. (Do not enter FmNT).
16. In Windows NT, open the HyperTerminal as follows:
a. Click on
Start > Programs > Accessories > HyperTerminal.
b. Click on the
HyperTerminal icon.
c. In the Connection Description dialog box, enter a name.
d. In the Connect To dialog box, select
Using COM2.
e. In the Com2 Properties dialog box, enter the settings shown in Table 8-44 and then click
OK.
Table 8-44 Com2 Settings
COM2 Properties
Bits per Second 9600
Data Bits 8
Parity None
Stop Bits 1
Flow Control None
No
Yes