Installation Manual
8-142 Troubleshooting
16. Type [P] then press [Enter] and observe the digital readout.
17. Does the Scale return a numeric value?
Go to Scale Communications Verifications in
this subsection.
Go to Step 18
18. Replace the Scale Cup and repeat Step 16.
19. Did the value change?
Scale (P/N 62-1614-00) has failed. See Parts
Replacement
section. (Balance and Control
Modules must be a matched set by serial number.)
Contact Asymtek Technical Support
20. Exit FmNT, shut down Windows NT, press the black OFF (0) button on the Operator’s Console,
switch Main Circuit Breaker to
OFF (0) position, and disconnect Power Cable from the facility outlet.
21. Check the Scale Interconnect Cable (P/N 194160) for continuity (Pinout is 1 to 1).
22. Is there continuity?
Scale Interconnect Cable (P/N 194160) has failed.
See
Parts Replacement section.
Scale (P/N 62-1614-00) has failed. See
Parts Replacement section. (Balance
and Control Modules must be a matched set by serial number)
No
Yes
No
Yes
No
Yes