Installation Manual
8-122 Troubleshooting
25. Check continuity (Pinout is 1 to 1) of Belt Motor Cable (Interconnect PWA to Motors).
26. Is there continuity?
Belt Motor Cable (P/N 06-4622-00) has failed.
See
Parts Replacement section.
Belt Motor (P/N 07-0716-00) has failed. See
Parts Replacement section.
Width Motor and Cable Verification
1. Exit FmNT, shut down Windows NT, press the black
OFF (0) button on the Operator’s Console,
switch Main Circuit Breaker to
OFF (0) position, and disconnect Power Cable from the facility outlet.
2. Remove the Conveyor Controller Cover.
3. Connect the Power Cable to the facility outlet, switch Main Circuit Breaker to
ON (I) position, press
the green
ON (I) button on the Operator’s Console, and enter FmNT.
4. While jogging the Width Motor, observe the six LEDs (labeled D32 to D36) located in the middle of
the Conveyor Controller Lower Board.
5. Did the six LEDs appear to illuminate intermittently as the Motor turns?
Conveyor Controller (P/N 62-1677-00 on X-1020, or
P/N 62-1676 on X-1010) has failed. See
Parts
Replacement
section.
Go to Step 6
6. Exit FmNT, shut down Windows NT, press the black OFF (0) button on the Operator’s Console,
switch Main Circuit Breaker to
OFF (0) position, and disconnect Power Cable from the facility outlet.
7. Check resistance between connector pins on Width Motor as specified in Table 8-38.
Table 8-38 Width Motor Pin Connections
Connector Pins
Between Pin 1 and Pin 2
Between Pin 3 and Pin 4
8. Is resistance 1.5 ohms between pins on the Width Motor Connector?
Width Motor (P/N 07-0716-00) has failed. See
Parts Replacement section.
Go to Step 9
No
Yes
No
Yes
No
Yes